Determination of Natamycin by High Performance Liquid Chromatography
LIU Yue,WANG Ya-li,ZHANG Hui-min,YU Lan,AN Lan-lan
NCPC New Drug Recearch and Development Corporation, Ltd., National Engineering Research Center of Microbial Medicine,Hebei Industry Microbial Engineering and Technology Research Center, Shijiazhuang 050015,China
An HPLC method for determination of natamycin was established. A C18 column(4.6 mm×250 mm,5 μm) was used with the phase of ammonium acetate-acetonitrile(760∶240), at the detection wavelength of 303 nm and the column temperature of 45 ℃.The calibration curves of natamycin were liner in the concentration rages of 0.04~0.3 mg/mL. The regression equation was y=52 196 216x + 202 922, r=1.000.The recovery was 99.73% with RSD 0.74%.The method was simple, accurate and reproducible.