| Uncertainty analysis of calibration for chlorine content by sulfur chloride analyzer |
| Wang Zhaolin, Hao Jingkun, Zhang Rui |
| Institute of Metrology of Hebei Province, Shijiazhuang 050000, China |
|
|
|
|
Abstract Sulfur chloride analyzer is widely used for determination of chlorine content in products in petroleum, petrochemical, quality inspection and other fields. The uncertainty of the measurement results directly affects product quality. The uncertainty of instrument indication error was analyzed. It mainly includes uncertainty introduced by measurement repeatability, instrument indication resolution, the micro-sampler, value of standard materials. As a result, the uncertainty of the chlorine content of the sulfur chloride analyzer was mainly derived from the uncertainty of the standard material.
|
|
|
|
|
|
| [ 1 ] JJF 1001 - 2011. 通用计量术语及定义[ J ]. 北京:中国计量出版社,2011.
[ 2 ] JJF 1059. 1-2012. 测量不确定度评定与表示[ S ]. |
|
|
|